ADC ML-190 Technical Information

Browse online or download Technical Information for Electric laundry dryers ADC ML-190. ADC ML-190 Technical information User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 1 ML 200 Version 2.2X The software version is visualised when power

Page 2

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 10 - Weld to the pipeline a pipe Ø 1". ATTENTION: verify "

Page 3

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 11 SIZE 1 from ND 80 up to ND 500SIZE 2SIZE 3from ND 80 up to ND 1000

Page 4 - INTRODUCTION

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 12 MS 3770 Assemble Instructions Flow directionZweld the jacket tothe

Page 5

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 13 ISOIL100,08,080,076,0170,0L126,093,054,0112,0133,0106,030,0146,0L

Page 6 - Recommended lifting

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 14 MS 1000 OVERALL DIMENSIONS ISOIL100,08,0126,093,054,0112,0133,0106

Page 7 - MS 500-1000-2400-2500

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 15 MS 2400 OVERALL DIMENSIONS ISOIL100,08,0100DHL126,093,054,0112,013

Page 8

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 16 MS 2500 PN10/64 OVERALL DIMENSIONS DLDNH10025PN 10L H D32L H D40L

Page 9 - Z=L-S-1/8D

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 17 MS 2500 ANSI 150-300 OVERALL DIMENSIONS ISOIL100,08,0126,093,054,

Page 10

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 18 2520013311215010810810884120133112108150133112150108108CONVERTER:

Page 11 - Z=L-S-1/8D-32

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 19TECHNICAL CHARACTERISTICS ELECTRIC CHARACTERISTICS Classification o

Page 12 - Flow direction

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 2 INDEX - Introduction…………………………………………………………………………………………………………………………

Page 13 - 3 6 10 15 20

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 20 OPERATING TEMPERATURE OF THE SENSOR SENSOR WITH EBONITE LINING SENSOR WITH PP LINING SENSOR

Page 14

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 21 REAR VIEW OF CONVERTER 1 2 3 4 5 6 7 8 9 1011 12 13 14 15 16 17 18

Page 15 - CHANGE WITHOUT NOTICE

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 22GROUNDING INSTRUCTION (with METALLIC PIPE) ALWAYS connect the senso

Page 16 - A - B - C

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 23For the sensors MS 500-1000-2500, in the cases listed below, see the following instruction:

Page 17

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 24 ELECTRICAL CONNECTIONS SENSOR-CONVERTER Sudden movements of the e

Page 18

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 25 CONVERTER POWER SUPPLY Terminal block M1 diagram IN/OUT - befor

Page 19 - TECHNICAL CHARACTERISTICS

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 26 Totaliser reset from remote input (if active functions: 30,31,32,3

Page 20

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 27(Out1) (Out2) (Out1) (Out2) On/off outputs wiring(Up to 1250 HZ) (low frequency): O

Page 21 - Terminal block M3

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 28WIRING TERMINAL M2 WITH MODULE: OUT 3 RS232 OUT 3

Page 22

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 29ACCESS TO THE INSTRUMENT The ML 200 programming keyboard is made up by three keys:

Page 23 - INSULATING GASKET

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 3 - Wiring terminal M2 with module……………….………………………………………………………………………

Page 24 - Coil cable

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 30 Sampling rate tasca trasparente per inserimento logo tasca trasp

Page 25 - 11 12 13 14 15 16 17 18 19 20

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 31 tasca trasparente per inserimento logo tasca trasparente per inser

Page 26 - Speed rate

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 32 MANUFACTURER PRE-SETTINGS The converter is programmed and delivere

Page 27

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 33 ACCESS CODES The information of this manual are related to all th

Page 28

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 34 If for several reasons you need to change the level of block of th

Page 29 - KEY BOARD

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 35ACCESS CONFIGURATION MENU From any visualizzation pages push this k

Page 30 - Sampling rate

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 36 FUNCTION N° MENU DESCRIPTION 1.SENSOR 1 ND 2 KA 3SET ND SET KA8 FS

Page 31

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 37 5 E.P. DETECT 6 AUTOZERO CAL. 7 E.P. CALIBR.ON/OFFSTART AUTOZERO

Page 32 - Dip switch

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 38FUNCTION MENU MENU 1.SENSOR (POS. 1) Nominal diameter Sensor [ND= XXXX] Senso

Page 33 - ACCESS CODES

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 39(POS. 3) Coefficient KL [KL= ±X.XXXX] Coefficient KL set. Leave the value at default (+0,000

Page 34 - BLOCK LEVELS

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 4 INTRODUCTION - This manual is integral part of the product. Read c

Page 35 - EXAMPLES

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 40(POS. 7) “empty pipe” calibration [E.P. CALIBR.] this function enables/disables the automatic

Page 36

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 41?? Metric units of measure cm3 0.001 Cubic centimetre ml 0.001 Millilitre l 1.000 Litre dm3

Page 37

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 42?? Metric mass units of measure G 0.001 Gram Kg 1.000 Kilogram T 1000.000 Ton ?? British

Page 38 - MENU 1.SENSOR

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 43(POS.10) Pulse value channel 1 and unit of measure of the totalizers [IMP1=dm³ X.XXXXX] Settin

Page 39

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 44(POS.13) Pulse duration channel 2 [TPUL2=ms XXXX.XX] Setting of the duration of the pulse ge

Page 40 - MENU 2.SCALES

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 45MENU 3.MEASURE (POS. 17) Time constant [TCONST=s XXXX.X] Time constant set. This parameter

Page 41

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 46(POS. 18) Acceleration threshold [SKIP THR=% XXX] Acceleration threshold set. The accelerati

Page 42

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 47(POS. 19) Peak cut off threshold [PEAK THR=% XXX] Anomalous signal pick c

Page 43 - IMP2=dm³ X.XXXXX]

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 48(POS. 22) Automatic scale change enable [AUTORANGE= ON/OFF] Enables/disables the automatic ch

Page 44

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 49MENU 4.ALLARS (POS. 24) Maximum flow rate threshold [MAX THR=% XXX] Maximum flow rate val

Page 45 - MENU 3.MEASURE

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 5 SAFETY Dangerous voltage, terminals identified by this symbol are su

Page 46

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 50The allowed range is from 0 to 120% of the 0..20 mA scale, 120% corresponds to 24 mA and does n

Page 47

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 51(POS. 31) Partial + totalizer reset enable [P+ RESET = ON/OFF] Partial direct (positive) flo

Page 48

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 52that the liquid is perfectly staying still. Even very small movement of the liquid may effect t

Page 49

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 53(POS. 41) Choice of the function corresponding to on/off output 3 [OUT4= XXXXXX]

Page 50 - MENU 5.INPUT

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 54(POS. 42) Choice of the function and the range of current output n.1[OUT.mA1=X÷XX ±]

Page 51

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 55 Example 1: out 4÷22 + Example 2: out 4÷20 –0+ I (mA) -100% full s

Page 52

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 56 (POS. 43) Choice of the function and the range of current output n.2 [OUTmA2= X÷XX ±] Choice

Page 53 -

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 57MENU 8.DISPLAY (POS.47) Totalisers number of decimals set [TOT.DECIMALS= X] Totalisers number

Page 54

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 58(POS. 53) Data logging time interval set [INTERV.(h)= X] Sampling time interval for the data

Page 55

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 59(POS. 58) Flow rate simulation enable [SIMULATION] Flow rate simulation enable. With

Page 56

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 6Flow direction GENERAL INFORMATION ON THE INSTALLATION RECOMMENDATION FOR MS 500-10

Page 57

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 60(POS. 61) Load factory pre-settings [LOAD FACT PRES.] Loads the pre-set programming of the fact

Page 58

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 61 Message Cause Action to take MAX ALARM The flow rate is higher than the maximum threshold s

Page 59 - MENU 11.INTERNAL DATA

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 62BATCH FUNCTION. ENABLE BATCH Enable one of the following functions to enable and program the

Page 60

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 63 ML 200: VISUALIZATION PAGE WITH BATCH FUNCTION IN PROGRESS Visualiz

Page 61

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 64N.B.: For each formula you can associate: Product quantity Product name Maximum time for batch

Page 62 - BATCH FUNCTION

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 65 IMPORTANT NOTES The start of the batch disables the following function: - R

Page 63

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 66 START UP AND MAINTENANCE OF THE INSTRUMENTS BEFORE STARTING UP TH

Page 64

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 67APPENDIX 1 TRASFORMATION CONVERTER FROM COMPACT VERSION TO SEPARATE VERSION

Page 65 - START/STOP BATCH:

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 68TRASFORMATION CONVERTER FROM SEPARATE VERSION TO COMPACT VERSION CoverFixing screwFixing plateO

Page 66

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 69APPENDIX 2 MEASURING RANGE ND min. 0….. 0,4 m/s max. 0….. 10 m/s 3 0………………………………10l/h 0……………

Page 67 - 1 2 3 4 5 6 7 8 9 10

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 7MS 500-1000-2400-2500 ISOILISOILLONG PIPEInstall the sensor away fro

Page 68

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 70 SENSOR NDPTFE LINING36101520MS 50036101519MS 240025324050658010023

Page 69

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 71APPENDIX 3 ERROR CURVE WITH CONVERTER ML 200 E(%)E(%)v (m/s)1 2 3 10

Page 70 - PTFE LINING

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 72 DECLARATION OF CONFORMITY according to ISO / IEC Guide 22 and EN 45014

Page 71 - 1 2 3 10

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 8MS 3700-3770 Install the sensor away from curves

Page 72 - DECLARATION OF CONFORMITY

C:\UFFICIO TECNICO\MANUALI\ML200\ML200 INGLESE\ 9 MS 3700 Sensor positioning BLOCK NUT Ø 16 MAFIXING NUT Ø 1" ME